Study on the measurement of the dynamic misalignment angle of the centrifuge

  • Zhong Chen*
  • , Feng Yuan
  • , Zhen Liang Ding
  • *Corresponding author for this work
  • Harbin Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

In the light of the problems in dynamic misalignment angle detection of the centrifuge, the misalignment angle measurement is presented. The main instrument is a dual-axis autocollimator. In the system, the vacuum optical path is used to shield the airflow to reduce the error due to air disturbance. The main links of the system are designed carefully and the errors are analyzed. Analysis and experiment show that the system uncertainty can be controlled within 0.2″.

Original languageEnglish
Pages (from-to)32-34
Number of pages3
JournalGuangxue Jishu/Optical Technique
Volume31
Issue number1
StatePublished - Jan 2005

Keywords

  • Misalignment angle
  • Optical path shield
  • Photo-electrical autocollimator

Fingerprint

Dive into the research topics of 'Study on the measurement of the dynamic misalignment angle of the centrifuge'. Together they form a unique fingerprint.