Abstract
A novel design of the rotor static balance measuring instrument structure is introduced, in which an agate blade bearing and autocollimator are adopted, and high accuracy measurement for rotor static balance is realized. A novel mechanism combining coarse fine double blades bearing ensures high precision and wide range, reduces the wear and tear of blades, and improves the efficiency of measurement. The sensitivity coefficient of the instrument is adjusted by measuring the swing period of rotor bearer. The instrument is calibrated by standard weights, experiment results show that the precision of the instrument is about 4 mg · mm and the measurement range is about 2000 mg · mm.
| Original language | English |
|---|---|
| Pages (from-to) | 258-261 |
| Number of pages | 4 |
| Journal | Jiliang Xuebao/Acta Metrologica Sinica |
| Volume | 32 |
| Issue number | 3 |
| DOIs | |
| State | Published - May 2011 |
Keywords
- Gyroscope rotor
- Inertial navigation
- Metrology
- Moment measurement
- Static balance measuring instrument
Fingerprint
Dive into the research topics of 'Study on static balance measurement of gyroscope'. Together they form a unique fingerprint.Cite this