Abstract
A simple method was proposed by using a tunable attenuator fitted in the reference or measurement arm of a heterodyne laser interferometer to adjust the values of mixing laser beams while the spectrum of the measurement signal is monitored using a signal analyzer. The effectiveness of the proposed method in reducing the first-order optical nonlinearity was verified through experiments. Results indicated that the peak value of the first-order optical nonlinearity could be reduced from 5.15 to 0.24 nm. It was therefore concluded that the proposed method was applicable to ultraprecision laser interferometry.
| Original language | English |
|---|---|
| Pages (from-to) | 6321-6326 |
| Number of pages | 6 |
| Journal | Applied Optics |
| Volume | 54 |
| Issue number | 20 |
| DOIs | |
| State | Published - 10 Jul 2015 |
Keywords
- And metrology
- Instrumentation
- Interferometry
- Measurement
- Metrological instrumentation
- Optical instruments
Fingerprint
Dive into the research topics of 'Simple method for reducing the first-order optical nonlinearity in a heterodyne laser interferometer'. Together they form a unique fingerprint.Cite this