Abstract
A novel R-wave detection algorithm of DCG signal is proposed. The algorithm takes the average two-way slope and relative height as its main characteristics. It can detect the R-wave fast and accurately. The feasibility of the algorithm is verified by MIT-BIH Long-Term ECG database and the Holter records from FAHHMU. Experimental results indicate that this algorithm has a much more precise detection rate(98.3%) than the maximum double-searching technology and the difference operation method(DOM)(95.2% and 90.7%).
| Original language | English |
|---|---|
| Pages (from-to) | 74-80 |
| Number of pages | 7 |
| Journal | Tianjin Daxue Xuebao (Ziran Kexue yu Gongcheng Jishu Ban)/Journal of Tianjin University Science and Technology |
| Volume | 47 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 2014 |
| Externally published | Yes |
Keywords
- Average two-way slope
- DCG signal
- MIT-BIH Long-Term ECG database
- Relative height