Skip to main navigation Skip to search Skip to main content

Note: Measuring instrument of singlet oxygen quantum yield in photodynamic effects

  • School of Electrical Engineering and Automation, Harbin Institute of Technology
  • School of Computer Science and Technology, Harbin Institute of Technology
  • Harbin Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

Using diphenylisobenzofuran (C20H14O) as a singlet oxygen (1O2) reporter, a comparison method, which can be used to measure the singlet oxygen quantum yield (ΦΔ) of the photosensitizer quantitatively, is presented in this paper. Based on this method, an automatic measuring instrument of singlet oxygen quantum yield is developed. The singlet oxygen quantum yield of the photosensitizer hermimether and aloe-emodin is measured. It is found that the measuring results are identical to the existing ones, which verifies the validity of the measuring instrument.

Original languageEnglish
Article number066102
JournalReview of Scientific Instruments
Volume88
Issue number6
DOIs
StatePublished - 1 Jun 2017

Cite this