Skip to main navigation Skip to search Skip to main content

Methods for measuring and controlling the movable coil position of the joule balance

  • National Institute of Metrology China
  • Harbin Institute of Technology

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

A system based on laser heterodyne interferometer is proposed in this paper to measure and control the movable coil position of a joule balance. A damping system is used to suppress the movement of movable coil in the horizontal direction while a piezoelectric ceramic control unit with PID controller is used to inhibit the vibration and long term drift of movable coil in the vertical direction. The effectiveness of the proposed method in measuring and controlling the movable coil position of a joule balance is proved through experiments. Experiment results indicate that the displacement of movable coil in the vertical direction can be reduced from 400 nm to 50 nm while the drift of movable coil is successfully inhibited.

Original languageEnglish
Title of host publicationCPEM 2014 - 29th Conference on Precision Electromagnetic Measurements, Digest
PublisherInstitute of Electrical and Electronics Engineers Inc.
Pages636-637
Number of pages2
ISBN (Electronic)9781479952052
DOIs
StatePublished - 12 Sep 2014
Event29th Conference on Precision Electromagnetic Measurements, CPEM 2014 - Rio de Janeiro, Brazil
Duration: 24 Aug 201429 Aug 2014

Publication series

NameCPEM Digest (Conference on Precision Electromagnetic Measurements)
ISSN (Print)0589-1485

Conference

Conference29th Conference on Precision Electromagnetic Measurements, CPEM 2014
Country/TerritoryBrazil
CityRio de Janeiro
Period24/08/1429/08/14

Fingerprint

Dive into the research topics of 'Methods for measuring and controlling the movable coil position of the joule balance'. Together they form a unique fingerprint.