Skip to main navigation Skip to search Skip to main content

Measuring error of phase modulation laser doppler shift measuring method

  • Jun Du
  • , Yancheng Qu*
  • , Weijiang Zhao
  • , Lijie Geng
  • , Chuang Liu
  • , Ruiliang Zhang
  • , Zhenlei Chen
  • *Corresponding author for this work
  • Harbin Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

By investigating distribution regularity of measuring error theoretically and experimentally, this paper builds measuring error mode of phase modulation laser Doppler shift measuring method based on its operating principle, which can be used to quantitative analyze and optimization design of Doppler shift measuring method for acquiring optimum measuring results in practical application. This error mode is utilized to theoretically investigate the influence of phase modulation depth and frequency on its frequency shift measuring accuracy and so on. It is found that measuring accuracy can reach the maximum when phase modulation frequency is equal to 0.63 times of full width of half maximum of Fabry-Perot interferometer transmission curve and phase modulation depth is 1.08.

Original languageEnglish
Article number0712001
JournalGuangxue Xuebao/Acta Optica Sinica
Volume34
Issue number7
DOIs
StatePublished - Jul 2014

Keywords

  • Beat frequency
  • Doppler shift
  • Phase-modulating
  • Physical optics
  • Remote sensing

Fingerprint

Dive into the research topics of 'Measuring error of phase modulation laser doppler shift measuring method'. Together they form a unique fingerprint.