Skip to main navigation Skip to search Skip to main content

Homodyne laser vibrometer modified by an LCVR for measurement at the nanometer level

  • Haijin Fu
  • , Xiong Xinkang
  • , Zheng Wang
  • , Pengcheng Hu*
  • , Ke Wang
  • , Jiubin Tan
  • *Corresponding author for this work
  • Ministry of Industry and Information Technology
  • Harbin Institute of Technology
  • National Institute of Metrology China

Research output: Contribution to journalArticlepeer-review

Abstract

The existence of periodic nonlinear error restricts the performance of the homodyne laser vibrometer in sub-fringe amplitude vibration measurement. A homodyne laser vibrometer with nanoscale-amplitude detectability by using a liquid crystal variable retarder (LCVR) is proposed. The LCVR introduces an extra variation of optical path difference larger than the laser wavelength to acquire a full ellipse so that the nonlinearity correction parameters could be pre-extracted. The experiments showed that the nonlinear error could be well suppressed with the correction process based on the pre-extracted parameters, and the detectable minimum amplitude is less than 1 nm. In addition, measurement of vibration with the reflectivity of measured targets down to 0.048% was achieved with an automatic-gain-control module.

Original languageEnglish
Pages (from-to)775-782
Number of pages8
JournalApplied Optics
Volume61
Issue number3
DOIs
StatePublished - 20 Jan 2022

Fingerprint

Dive into the research topics of 'Homodyne laser vibrometer modified by an LCVR for measurement at the nanometer level'. Together they form a unique fingerprint.

Cite this