Skip to main navigation Skip to search Skip to main content

Determination of wear metals in aviation lubricating oil by using a simultaneous MPT spectrometer with an automatic reflux extraction system

  • Shu Hua Wang*
  • , Guo Dong Feng
  • , Xiang Yu Zou
  • , Jie Jiang
  • , Yan Fu Huan
  • , Wen Chao Dou
  • , Qin Han Jin
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

A novel, simple and economic sample pretreatment system based on reflux extraction was developed for the determination of wear metals in the used aviation lubricating oil by using a simultaneous microwave plasma torch atomic emission spectrometer (MPT-AES). The operating parameters for both sample pretreatment and analysis were studied and optimized. The detection limits are 0.084, 0.043, 0.022, 0.052, 0.064 μg/g for Zn, Cd, Cu, Fe and Pb, respectively. The relative standard deviations are less than 5% (n=5), and the recovery ratios of these elements are in the range of 94.5%-107.4%. The new system was successfully applied to the analysis of the lubricating oil samples.

Original languageEnglish
Pages (from-to)2048-2051
Number of pages4
JournalGaodeng Xuexiao Huaxue Xuebao/Chemical Journal of Chinese Universities
Volume27
Issue number11
StatePublished - Nov 2006
Externally publishedYes

Keywords

  • Aviation lubricating oil
  • Reflux extraction
  • Simultaneous MPT-AES
  • Wear metal

Fingerprint

Dive into the research topics of 'Determination of wear metals in aviation lubricating oil by using a simultaneous MPT spectrometer with an automatic reflux extraction system'. Together they form a unique fingerprint.

Cite this