A study on the minimum detectable energy density precise measurement method of STIL

  • J. Wei
  • , Q. Wang
  • , J. Sun
  • , J. Gao
  • Harbin Institute of Technology

Research output: Contribution to journalArticlepeer-review

Abstract

This paper presents a precise measurement method of streak tube imaging LIDAR's (STIL's) minimum detectable energy density and the measurement result of a K008 streak tube. A Gaussian expression was used to approximate the one-dimensional intensity distribution of the laser spot. The laser pulse energy was measured with a sensitive photon counter while the diameter of the laser spot was measured by the single knife-edge method. After determining the parameters using the Gaussian expression and calculations, the authors concluded that the minimum detectable energy density of STIL in this experiment was 5.27 ± 0.11 × 108 J mm-2.

Original languageEnglish
Article number045302
JournalLaser Physics
Volume23
Issue number4
DOIs
StatePublished - Apr 2013